Phase | Description | Applications | Bio-Rad® Aminex® | Supelco® Supelcogel™ | Waters® Sugar-Pak™ | Transgenomic® CARBOSep™ | Sepax® Carbomi® |
---|---|---|---|---|---|---|---|
RPM-Monosaccharide (USP L34 packing) | 8 % cross-linked LEAD | – 100 x 7.8 mm dimensions available for sugar alcohol analysis according to the USP procedure – Monosaccharides and sugar alcohol analysis – Cellobiose, glucose, xylose, arabinose, and mannose and other cellulose products |
HPX-87P 125-0098 |
Supelcogel Pb |
N/A | CARBOSep COREGEL-87P |
Carbomix Pb |
Counter Ion Analyte | RPM Pb2+ |
---|---|
Adonitol (Ribitol) | 20.15 |
D-Altrose | 15.82 |
D-(-)-Arabinose | 16.47 |
D-(+)-Cellobiose | 11.00 |
D-(+)-Digitoxose | 15.32 |
Dulcitol | 33.25 |
Meso-Erythritol | 19.82 |
D-(-)-Fructose | 17.71 |
L-(-)-Fucose | 16.19 |
D-(+)-Galactose | 14.94 |
Gentiobiose | 10.53 |
D-(+)-Glucose | 14.94 |
Inositol | 18.87 |
Isomaltose | 11.28 |
Lactose | 11.89 |
Lactulose | 13.95 |
D-Lyxose | 16.66 |
Counter Ion Analyte | RPM Pb2+ |
---|---|
D-Maltose | 11.59 |
Maltotriose | 11.02 |
Maltulose | 12.40 |
D-Mannitol | 24.90 |
D-(+)-Mannose | 16.39 |
Melibiose | 11.97 |
D-(+)-Melezitose | 9.94 |
D-(+)-Raffinose | 10.28 |
L-(+)-Rhamnose | 14.47 |
D-(-)-Ribose | 33.48 |
Salicin | 26.81 |
D-Sorbitol | 35.97 |
Stachyose | 9.72 |
Sucrose | 11.00 |
Trehalose | 11.01 |
Xylitol | 32.38 |
D-(+)-Xylose | 13.84 |
Column:
Rezex RPM-Monosaccharide
Bio-Rad® Aminex® HPX-87P |
|
Dimension:
300 x 7.8 mm
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
80 °C
|
|
Detection:
ELSD
|
|
Sample:
1. Glucose
2. Erythritol 3. Mannitol 4. Xylitol 5. Sorbitol |
Column:
Rezex RPM-Monosaccharide
Aminex HPX-87P |
|
Dimension:
300 x 7.8 mm
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
75 °C
|
|
Detection:
ELSD
|
|
Sample:
1. Melezitose
2. Maltose 3. Glucose 4. Mannose 5. Fructose 6. Ribitol |
Column:
Rezex RPM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0135-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
85 °C
|
|
Detection:
ELSD
|
|
Sample:
1. D-Cellobiose
2. α-D-glucose 3. D-(+)-Xylose 4. D-(+)-Galactose 5. D-(-)-Arabinose |
Column:
Rezex RPM-Monosaccharide
|
|
Dimension:
300 x 7.8 mm
|
|
Part No.:
00H-0135-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
75 °C
|
|
Detection:
RI
|
|
Sample:
1. Stachyose
2. Maltose 3. Glucose 4. Xylose 5. Galactose 6. Fructose 7. meso-Erythritol |
8. Mannitol 9. Salicin 10. Xylitol 11. Sorbitol |
Column:
Rezex RPM-Monosaccharide
|
|
Dimension:
100 x 7.8 mm
|
|
Part No.:
00D-0135-K0
|
|
Mobile Phase:
Water
|
|
Flow Rate:
0.6 mL/min
|
|
Temperature:
80 °C
|
|
Detection:
ELSD (Ambient)
|
|
Sample:
1. Mannitol
2. Sorbitol |
Compound | App ID No. |
---|---|
D-Cellobiose | 14967 |
D-(+)-Galactose | 14967 |
D-(+)-Xylose | 14967 |
D-(+)-Arabinose | 14967 |
Erythritol | 14968 |
Fructose | 5508, 14985 |
Galactose | 5508 |
Glucose | 5508, 14985 |
Maltose | 5508, 14985 |
Mannitol | 5508, 14968 |
Compound | App ID No. |
---|---|
Mannose | 14985 |
Melezitose | 14985 |
meso-Erythritol | 5508 |
Ribitol | 14985 |
Salacin | 5508 |
Sorbitol | 5508, 14968 |
Stachyose | 5508 |
Xylitol | 5508, 14968 |
Xylose | 5508 |
Rezex RPM-Monosaccharide |   |
---|---|
Part No. | 00H-0135-K0 |
Ionic Form | Lead |
Standard Dimensions | 300 x 7.8 mm |
Matrix | Sulfonated Styrene Divinyl Benzene |
Cross Linking | 8 % |
Particle Size (μm) | 8 |
Min. Efficiency (p/m) based on last peak | 35,000 |
Typical Pressure (psi @ Testing Flow Rate) | 190 |
Max. Pressure (psi @ Max Flow Rate) | 1,000 |
Max. Flow Rate (mL/min) | 1.0 |
Max. Temperature (°C) | 85 |
Typical Mobile Phase | Water |
pH Range | Neutral |
Guard Column Part | 03B-0135-K0 |
Cleaning, Regeneration and Storage |   |
---|---|
Inorganic Modifiers | 5 % Lead Nitrile |
Organic Modifiers (Max) | 5% Methanol, IPA, EtOH |
Avoid | Acids, Bases, Non-Lead Salts or Metal Ions, > 30 % Acetonitrile |
Cleaning Solvent | 100 % Water |
Flow Rate(mL/min) | 0.4 |
Temperature (°C) | 85 |
Duration (hrs) | 12 |
Regeneration Solvent | 0.1 M Pb(NO3)2 |
Flow Rate (mL/min) | 0.2 |
Duration (hrs) | 4-16 |
Ship/Storage Solvent | Water |