SPE HILIC stationary phase that excels at retention and recovery labeled released N-glycans. Microelution format allows for streamlined processing and clean-up of small sample volumes.
Brand bioZen |
Recommended Use Extraction and clean-up of N-glycans |
Reliable recovery and faster results
STEP 1 |
STEP 2 |
STEP 3 |
STEP 4 |
STEP 5 |
Start with Intact Glycoprotein | Deglycosylation Released N-Glycans by PNGase F in buffer solution |
Labeling Glycans 2-AA or 2-AB for accurate and sensitive glycan detection |
Clean-Up Removal of labeling reagent, salts, or surfactants |
Analysis |
96-Well Plate:
Biozen N-Glycan Clean-up
|
Part No.:
8M-S009-NGA
|
Condition:
200 µL Water
|
Equilibrate:
200 µL Water/Acetonitrile (15:85)
|
Load:
Labeled Sample (up to 400 µL of labeled sample)
|
Wash:
2x 600 µL Formic acid/Water/Acetonitrile (1:9:90)
|
Elute:
3x 30 µL 200 mM Ammonium acetate in Acetonitrile/Water (5:95) of labeled sample)
|
Dilute:
100 µL Dimethylformamide and then 210 µL of Acetonitrile
|
Biozen Columns (mm) | Biocompatible Guard Cartridges | |||
100 x 2.1 | 150 x 2.1 | for 2.1 mm | holder | |
Biozen 2.6 μm Glycan |
00D-4773-AN | 00F-4773-AN | AJ0-9800 | AJ0-9000 |
Glycan | Sum of Peak Area (%) | Multiplication Factor | Charge Number Share |
Asialo | 0.28 | 0 | 0 |
Monosialo- | 13.41 | 1 | 13.41 |
Disialo- | 4.44 | 2 | 8.88 |
Trisialo- | 52.49 | 3 | 157.47 |
Tetrasialo- | 29.4 | 4 | 117.6 |
Total peak area | 100% | Z-value= 297.36 |