For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Product Information
Phase Detail
Biozen N-Glycan Clean-Up<br class="t-last-br" />
Biozen Glycan sample preparation products offer targeted selectivity for N-glycan

Biozen N-Glycan Clean-Up

Easy Clean-up of Labeled, Released N-Glycans Using Solid Phase Extraction (SPE)

SPE HILIC stationary phase that excels at retention and recovery labeled released N-glycans. Microelution format allows for streamlined processing and clean-up of small sample volumes.

  • Small volume format allows for more concentrated samples
  • Minimal elution solvent
  • Excellent retention and recovery of labeled, released N-glycans
  • Faster clean-up methods to save on sample prep time

Brand
bioZen
Recommended Use
Extraction and clean-up of N-glycans

Protocol

N-Glycan Clean-up

Reliable recovery and faster results


STEP 1

STEP 2

STEP 3

STEP 4

STEP 5
Start with Intact Glycoprotein Deglycosylation
Released N-Glycans by PNGase F in buffer solution
Labeling Glycans
2-AA or 2-AB for accurate and sensitive glycan detection 
Clean-Up
Removal of labeling reagent, salts, or surfactants
Analysis

96-Well Plate:
Biozen N-Glycan Clean-up
Part No.:
8M-S009-NGA
Condition:
200 µL Water
Equilibrate:
200 µL Water/Acetonitrile (15:85)
Load:
Labeled Sample (up to 400 µL of labeled sample)
Wash:
2x 600 µL Formic acid/Water/Acetonitrile (1:9:90)
Elute:
3x 30 µL 200 mM Ammonium acetate in Acetonitrile/Water (5:95) of labeled sample)
Dilute:
100 µL Dimethylformamide and then 210 µL of Acetonitrile

bioZen LC Columns

Biozen Columns (mm) Biocompatible Guard Cartridges
  100 x 2.1 150 x 2.1 for 2.1 mm holder
Biozen 2.6 μm
Glycan
00D-4773-AN 00F-4773-AN AJ0-9800 AJ0-9000

Labeled Glycans from AGP


Sum of Peak Areas for Z-value Calculation

Glycan Sum of Peak Area (%) Multiplication Factor Charge Number Share
Asialo 0.28 0 0
Monosialo- 13.41 1 13.41
Disialo- 4.44 2 8.88
Trisialo- 52.49 3 157.47
Tetrasialo- 29.4 4 117.6
Total peak area 100% Z-value= 297.36