FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Improving Resolution for Size Exclusion Chromatography Methods by Optimization of Linear Velocity
Flow rate is a common method parameter for most adsorptive LC methods. However, SEC is a unique separation modality in that lower flow rates often yield improvements in separation and resolution can improve quite considerably simply by running at a lower linear velocity. This application note demonstrates the utility of method improvements by running at a lower flow rate using the Biozen dSEC-2 SEC column for the aggregate analysis of NIST mAb, a common standard protein.
Improving Resolution for Size Exclusion Chromatography Methods by Optimization of Linear Velocity
Flow rate is a common method parameter for most adsorptive LC methods. However, SEC is a unique separation modality in that lower flow rates often yield improvements in separation and resolution can improve quite considerably simply by running at a lower linear velocity. This application note demonstrates the utility of method improvements by running at a lower flow rate using the Biozen dSEC-2 SEC column for the aggregate analysis of NIST mAb, a common standard protein.