1C16:0 |
---|
2C18:1 cis 9 |
3C18:2 cis 9,12 |
4C18:3 cis 9,12,15 |
1C16:0 |
---|
2C18:1 cis 9 |
3C18:2 cis 9,12 |
4C18:3 cis 9,12,15 |
Column: |
Part No: |
Phase Name: ZB-FAME |
Sample Notes: Recommended Accessories:
Recommended Liner: Zebron PLUS Single Taper Z-Liner, 4 mm ID
Liner Part No.: AG2-0A13-05 (for Agilent systems)
Inlet Seal: AG0-8620 (Gold Plated Easy Seal)
Septum: AG0-4696 (PhenoRed-400)
Sample Preparation:
1. Weigh out 25 mg of oil into a 5 mL screw cap glass vial
2. Add 2 mL of BF3-methanol to sample vial
3. Heat vial at 60 °C for 15 minutes with periodic vortexing
4. Remove from heat and allow to attain room temperature
5. Add 1 mL water followed by 1 mL hexane
6. Vortex for 1 minute
7. Remove upper organic layer
8. Dry hexane layer over anhydrous sodium sulfate
9. Transfer to autosampler vial and inject on GC |