For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA 8081 by GC/ECD on ZB-CLPesticides-1 (Helium Gas)
22631
Separation Mode: Application Specific
Application Specific
EPA 8081
ZB-CLPesticides-1
GC
Insecticides
Environmental

EPA 8081 by GC/ECD on ZB-CLPesticides-1 (Helium Gas)

Analytes
12,4,5,6-TCMX (Surr)
2a-BHC
3g-BHC
4b-BHC
5d-BHC
6Heptachlor
7Aldrin
8Heptachlor epoxide
9trans-Chlordane
10cis-Chlordane
11Endosulfan I
124,4--DDE
13Dieldrin
14Endrin
154,4'-DDD
16Endosulfan II
174,4'-DDT
18Endrin aldehyde
19Endosulfan sulfate
20Methoxychlor
21Endrin ketone
22Decachlorobiphenyl (Surr)
Details
Official Method: EPA 8081
GC Conditions (App ID: 22631)
Column:
Part No:
Phase Name:
ZB-CLPesticides-1
Sample Notes:
Analytes are 250 ppb in hexane
Similar Applications
search
No Data