FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Ibuprofen by HPLC on Luna 5um C18(2) 250x4.6mm
19612
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Anti-Inflammatory Agents, Non-Steroidal
Clinical
Pharmaceutical/Biopharmaceutical
Pharmaceutical
Therapeutic / Clinical

Ibuprofen by HPLC on Luna 5um C18(2) 250x4.6mm

Analytes
1Ibuprofen
2Valerophenone
3Ibuprofen related compound C
Details
LC Conditions (App ID: 19612)
Column:
Brand Name:
Luna
Part No:
Phase Name:
C18(2)
Sample Note:
The USP Resolution requirement between any two peaks is not less than 2.5, and the tailing factor for any individual peak is not more than 2.5. On the LUNA 5um C18(2) the resolution between Ibuprofen and Valerophenone is 7.55; valerophenone and Ibuprofen related compound C is 6.11. The tailing factors for Ibuprofen, Valerophenone and Ibuprofen related compound C are 1.2, 1.45, and 1.26, respectively. The max back pressure was 215 bar.