FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Calibration Curve for Cyanuric acid Extraction (20-2000ng/mL) from Strata Melamine on Luna HILIC
18468
Separation Mode: Hydrophilic Interaction
Hydrophilic Interaction
HILIC
HPLC
Pollutant / Contaminant
Environmental / Health

Calibration Curve for Cyanuric acid Extraction (20-2000ng/mL) from Strata Melamine on Luna HILIC

Analytes
1Cyanuric acid
2Cyanuric-13C3
Details
LC Conditions (App ID: 18468)
Column:
Brand Name:
Luna
Part No:
Phase Name:
HILIC
Sample Note:
LC/MS/MS Condition: Column: Luna 3µm HILIC, 100x2.0mm; Flow rate: 0.4ml/min Mobile Phase: Isocratic; 90/10 ACN/100mM NH4 Formate pH 3.2; MS Detector: API 3000 LC/MS/MS with ESI (TurboIonSpray®) (www.Appliedbiosystems.com), Switch from -ve Ion Mode (for cyanuric) to Positive Ion (for Melamine); MRM; heater gas flow 7000 cc/min; heater temperature 450 °C Cyanuric acid QC - 200 ng % Accuracy 100.1% % CV 17% Calibration Curve of Extracted Samples conc. (ng) % Accuracy 20 104% 40 98% 200 95% 400 103% 2000 99.91%