For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Organochloropesticides (OCP) and Polychlorinated Biphenyls (PCB) in Milk and Milk Products by GC/ECD
17909
Separation Mode: Application Specific
Application Specific
ZB-MultiResidue-1
GC
Insecticides
Food & Beverage
Pesticide / Insecticide
Environmental / Health

Organochloropesticides (OCP) and Polychlorinated Biphenyls (PCB) in Milk and Milk Products by GC/ECD

Analytes
1alpha-HCH
2Hexachlorobenzene
3gamma-HCH
4beta-HCH
5delta-HCH
6PCB 28
7PCB 52
8Aldrin
9Heptachloro epoxide
102,4'-DDE (o,p-)
11PCB 101
124,4'-DDE (p,p-)
13Nitrofen
14Dieldrin
152,4'-DDD (o,p-)
16Endrin
174,4'-DDD (p,p-)
18PCB 153
19PCB 138
20PCB 180
21Mirex
Details
GC Conditions (App ID: 17909)
Column:
Part No:
Phase Name:
ZB-MultiResidue-1
System:
Agilent (HP) 6890
Sample Notes:
The application was kindly provided by the MLF, Dr.-Oskar-Farny-Institute in Wangen, Germany. Injector Gerstel-KAS3. Program: 50 °C with 12 °C/s to 300 °C, 3 min hold Solvent Venting Purge Time: 0.15 min, Splitless Time: 1 min, Equilibration Time 0.5 min, Initial Time: 0.15 min. Concentration: 25 ppb each for OCP + PCB, 100 ppb for Nitrofen
Similar Applications
search
No Data