For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
IgG_Dog (Native) Jupiter 300 C4
14908
Separation Mode: Reversed Phase
Reversed Phase
C4
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Protein
Biochemical Compound / Nutrient

IgG_Dog (Native) Jupiter 300 C4

Analytes
1IgG
Details
LC Conditions (App ID: 14908)
Column:
Brand Name:
Jupiter
Part No:
Phase Name:
C4
Sample Note:
These applications demonstrate the utility of using low amounts of isopropanol in the organic mobile phase to improve peak shape of hydrophobic proteins. App ID#14918 shows the elution of Ig-G on a Jupiter 300 C4 column where Acetonitrile is used as the organic mobile phase. App ID#14908 shows the same Ig-G run on the same column using an organic mobile phase that contains 20% isopropanol. Note the dramatic improvement in both recovery and peak shape for the Ig-G peak. The addition of isopropanol to the organic mobile phase improves the efficiency as well as the peak shape for large hydrophobic proteins due much in part to the increased hydrophobicity of the organic mobile phase upon addition of isopropanol. Especially when performing separations of Ig-G and other immunoglobin one should start with the use of a wide pore C4 column and look at using significant amounts of IPA as part of the organic mobile phase.
Similar Applications
search
No Data