For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Product Information
Phase Detail
Strata-X-Drug B and Strata-X-Drug B Plus SPE Products
Strata-X-Drug-B solid phase extraction (SPE) sorbent is a polymeric specialty phase that can be used to cleanup basic drugs of abuse

Strata-X-Drug B and Strata-X-Drug B Plus SPE Products

Quick and In-Well Hydrolysis Solid Phase Extraction (SPE)

Proprietary sorbent that is designed and quality controlled for basic drugs of abuse analysis. This sorbent does not require a conditioning/equilibration step to save time and help improve throughput. With Strata-X-Drug B Plus, the hydrolysis step is performed in the well, providing even more time savings and reducing transfer steps.

  • QC Tested for Drug of Abuse
  • No Conditioning or Equilibration Steps Required

Brand
Strata-X
Recommended Use
Extraction of basic drugs from biological samples

General Extraction

  • Extract over 11 drugs using the same sorbent
  • Consolidate methods; 3 methods is all you’ll need to extract 6 different drug classes
  • Time and solvent savings; no conditioning required
  • Consistent, sensitive, and high recoveries of SAMHSA regulated drugs

Retention Mechanism

Strata-X Drug B Protocol

 

Strata-X Drug B Plus Protocol

Starting Method

General Urine Pre-treatment Recommendation

  1. Combine urine sample with appropriate buffer.
  2. Add β-Glucuronidase enzyme according to manufacturer’s recommendation (If using Strata-X-Drug B Plus, these steps can be performed in-well).
  3. Incubate for specified amount of time (If using Strata-X-Drug B Plus,     pulse vacuum to load solution onto sorbent).
  4. Follow appropriate SPE protocol based on analytes of interest.

Choose Your Analytes

  Opiates, 6-MAM, PCP, Amphetamines, Propoxyphene, Healthcare Opiates,
and Methadone*
Marijuana Metabolites Cocaine Metabolites
Condition Not Required Not Required Not Required
Load Pre-treated  urine sample Pre-treated  urine sample Pre-treated  urine sample
Wash 1 600 µL of 100 mM Sodium acetate buffer (pH 5.0) 600 µL of 100 mM Sodium acetate buffer (pH 5.0) 600 µL of 0.1 N Hydrochloric acid
Wash 2 600 µL Methanol 600 µL of Acetonitrile/100 mM Sodium acetate buffer (pH 5.0) (30:70) 600 µL Methanol
Dry 10 minutes under full vacuum 15 minutes under full vacuum 10 minutes under full vacuum
Elute 2x 300 µL of Ethyl acetate/Isopropanol/
Ammonium hydroxide (70:20:10)
2x 300 µL of Ethyl acetate/Isopropanol
(85:15)
2x 300 µL of Ethyl acetate/Isopropanol/
Ammonium hydroxide (70:20:10)



* Opiates, 6-MAM, PCP, Amphetamines, Propoxyphene, Healthcare Opiates, and Methadone can be extracted simultaneously or separately using the same SPE methodology. Methods are written for 30 mg/well Strata-X-Drug B 96-well plate; however they can be scaled to accommodate smaller or larger sample sizes and sorbent masses.