FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Doxepin HCl - USP Method
9230
Separation Mode: Reversed Phase
Reversed Phase
C8(2)
HPLC
Antipruritics
Pharmaceutical
Therapeutic / Clinical

Applications

Doxepin HCl - USP Method
Analytes
1Doxepin
2Doxepin
Compound Name
Doxepin
CID
667477
Molecular Formula
C19H21NO
Molecular Weight
279
No. Hydrogen Bond Acceptors
2
Details
LC Conditions (App ID: 9230)
Column:
Brand Name:
Part No:
Phase Name:
C8(2)
Elution Type:
Isocratic
Mobile Phase:
A: 61:39 200mM Monobasic Sodium Phosphate:Methanol, adjusted to pH 2.5
Gradient Profile:
Step No.Time(min)%A
10100
Flow Rate:
1 mL/min
Temperature:
50°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 254 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C8 4 x 3.0mm, 10/Pk Part No.: AJ0-4290 - Holder Part No.: KJ0-4282