For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Fast Opiates Panel on Kinetex 2.6µm Phenyl-Hexyl, 50x3mm
23740
Separation Mode: Reversed Phase
Reversed Phase
Phenyl-Hexyl
HPLC
Analgesics, Opioid

Applications

Fast Opiates Panel on Kinetex 2.6µm Phenyl-Hexyl, 50x3mm
Analytes
16-MAM
2Buprenorphine
3Codeine
4EDDP
5Fentanyl
6Gabapentin
7Hydrodcodone
8Hydromorphone
9Meperidine
10Methadone
11Morphine
12Naloxone
13Norbuprenorphine
14Norfentanyl
15Norhydrocodone
16Noroxycodone
17O-Desmethyltramadol
18Oxycodone
19Oxymorphone
20Pregabalin
21Tramadol
Compound Name
6-Acetylmorphine
CID
5462507
Molecular Formula
C19H21NO4
Molecular Weight
327
No. Hydrogen Bond Acceptors
5
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 23740)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Phenyl-Hexyl
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic acid in water
B: 0.1% Formic acid in methanol
Gradient Profile:
Step No.Time(min)%A%B
109010
20.29010
31.56040
42595
52.5595
62.519010
749010
Flow Rate:
0.7 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (600°C)
Detector Info:
SCIEX Triple Quad™ 4500 System

Order Items Used in This Application

Similar Applications
search
No Data