For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Curcuminoids on Kinetex 2.6u C18 150x4.6mm
22914
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Antineoplastic Agents

Applications

Curcuminoids on Kinetex 2.6u C18 150x4.6mm
Analytes
1Bis-demethoxy-curcumin
2Demethoxy-curcumin
3Curcumin
Compound Name
Bisdemethoxycurcumin
CID
5315472
Molecular Formula
C19H16O4
Molecular Weight
308
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 22914)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% Formic Acid in Water
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
104555
283070
390100
4100100
5114555
6134555
Flow Rate:
1 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 425 nm (22°C)