FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Vitamin B1 and B6 chromatogram of LLOQ in Human Blood sample
22794
Separation Mode: Reversed Phase
Reversed Phase
C18(2)
HPLC
Vitamin B Complex

Applications

Vitamin B1 and B6 chromatogram of LLOQ in Human Blood sample
Analytes
1Pyridoxal 5-phosphate (PLP)
2Thiamin Diphosphate (TDP)

Analyte not found

Details
LC Conditions (App ID: 22794)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: 25 mM Na2HPO4 (pH7.0)
B: MeOH
Gradient Profile:
Step No.Time(min)%A%B
10.01955
21.88515
33.55050
44.55050
54.51955
66.5955
Flow Rate:
600 mL/min
Temperature:
28°C
Detection:
Fluorescence (FLUOR) (22°C)
Detector Info:
Shimadzu RF-20A

Order Items Used in This Application

Similar Applications
search
No Data