FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Mycotoxins extracted from corn using roQ by LC/MS/MS (Positive mode)
22711
Separation Mode: Reversed Phase
Reversed Phase
XB-C18
HPLC
Carcinogens, Environmental

Applications

Mycotoxins extracted from corn using roQ by LC/MS/MS (Positive mode)
Analytes
115-Acetyldeoxynivalenol
2Aflatoxin G2
3Aflatoxin G1
4Unknown Interference
5Aflatoxin B2
6Aflatoxin B1
7Diacetoxyscirpenol
8HT2-Toxin
9Fuminosin B1
10T2-Toxin
11Ochratoxin A
12Fuminosin B2

Analyte not found

Details
LC Conditions (App ID: 22711)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
XB-C18
Elution Type:
Gradient
Mobile Phase:
A: 5 mM ammonium acetate w/ 0.5% acetic acid
B: 5 mM ammonium acetate in methanol w/ 0.5% acetic acid
Gradient Profile:
Step No.Time(min)%A%B
10955
22955
352080
45.2298
58298
Flow Rate:
0.45 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
API 5000
Sample Note:
roQ extraction: 1) 5 g of corn is extracted using KS0-8909 with 10 mL of water and 10 mL of ACN with 1.0 % formic acid. 2. KSO-8924 is used for dSPE clean up.

Order Items Used in This Application

Similar Applications
search
No Data