For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Full Pain Panel using Kinetex 2.6u Biphenyl 50x2.1mm
22122
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Antiemetics
Clinical
Pharmaceutical/Biopharmaceutical
Forensics/Toxicology
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical
Therapeutic / Clinical

Applications

Full Pain Panel using Kinetex 2.6u Biphenyl 50x2.1mm
Analytes
1Morphine
2Oxymorphone
3Hydromorphone
4Amphetamine
5Methamphetamine
6MDA
7Naloxone
8Codeine
96-MAM
10Oxycodone
11MDMA
12Naltrexone
13Hydrocodone
14MDEA
15Norfentanyl
16Tramadol
17Benzoylecgonine
18Meprobamate
19Normeperidine
20Meperidine
21Norbuprenorphine
22Fentanyl
23Buprenorphine
24Flurazepam
25PCP
26Carisoprodol
27Propoxyphene
28Midazolam
29Sufentanil
30Norpropoxyphene
31EDDP
32Lorazepam
33Methadone
34Clonazepam
35Oxazepam
36Hydroxyalprazolam
37Nordiazepam
38Flunitrazepam
39Temazepam
40Alprazolam
41Diazepam
Compound Name
Morphine
CID
5288826
Molecular Formula
C17H19NO3
Molecular Weight
285
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
2
Details
LC Conditions (App ID: 22122)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Biphenyl
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.1% formic acid
B: Methanol with 0.1% formic acid
Gradient Profile:
Step No.Time(min)%A%B
109010
20.59010
327525
44.52080
54.511585
65.51585
75.519010
879010
Flow Rate:
0.6 mL/min
Temperature:
40°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm C18 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase
Kinetex 2.6 µm F5 100 Å, LC Column 50 x 4.6 mm, Ea
Reversed Phase