FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
ASTM D6352 on Zebron ZB-1XT SimDist
22015
Separation Mode: Application Specific
Application Specific
ASTM D 6352
ZB-1XT SimDist
GC
Fuels

Applications

ASTM D6352 on Zebron ZB-1XT SimDist
Analytes
1C9
2C10
3C11
4C12
5C20
6C36
7C38
8C40
9C60
10C62
11C64
12C76
13C78
14C80

Analyte not found

Details
Official Method: ASTM D 6352
GC Conditions (App ID: 22015)
Column:
Part No:
Phase Name:
ZB-1XT SimDist
Injection:
On-Column @ 43°C, 1.0 μL
Oven Program:
30 C to 410 C at 10 C/min hold for 2 min
Carrier Gas:
Constant Flow Helium, 21.9 mL/min
Detection:
Flame Ionization (FID) (450°C)
Sample Notes:
Customer sample @ 0.5% in CS2

Order Items Used in This Application

Similar Applications
search
Zebron ZB-DHA-PONA, GC Cap. Column 100m x 0.25mm x 0.5µm, Ea
Application Specific
Zebron ZB-1XT SimDist Metal, GC Cap. Column 10 m x 0.53 mm x 2.65 µm, Ea
Application Specific
Zebron ZB-1XT SimDist Metal, GC Cap. Column 5 m x 0.53 mm x 0.15 µm, Ea
Application Specific
Zebron ZB-1XT SimDist Metal, GC Cap. Column 5 m x 0.53 mm x 0.15 µm, Ea
Application Specific
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
Application Specific
Zebron ZB-MultiResidue-2, GC Cap. Column 30 m x 0.25 mm x 0.20 µm, Ea
Application Specific
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.18 µm, Ea
Standard