For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Mycotoxin analysis using Kinetex 2.6u Biphenyl 50x2.1mm
21951
Separation Mode: Reversed Phase
Reversed Phase
Biphenyl
HPLC
Poisons
Food & Beverage
Environmental
Agriculture
Product / Trade Good
Environmental / Health

Applications

Mycotoxin analysis using Kinetex 2.6u Biphenyl 50x2.1mm
Analytes
1 15-Acetyldeoxynivalenol
2 DAS
3 FB1
4 HT2 Toxin
5 FB2
6 T2 Toxin
7 Aflatoxin M1
8 Aflatoxin G2
9 Ochratoxin A
10 Aflatoxin G1
11 Aflatoxin B2
12 Aflatoxin B1

Analyte not found

Details
LC Conditions (App ID: 21951)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
Biphenyl
Elution Type:
Gradient
Mobile Phase:
A: 5mM ammonium acetate with 0.1% acetic acid
B: Methanol with 5mM ammonium acetate with 0.1% acetic acid
Gradient Profile:
Step No.Time(min)%A%B
10982
22982
352080
45.2298
58298
68.01982
711982
Flow Rate:
0.45 mL/min
Temperature:
22°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 4000™

Order Items Used in This Application

Similar Applications
search
No Data