For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Immunosuppressants by LCMS using Luna 5um Phenyl-Hexyl 50x2.1mm
20688
Separation Mode: Reversed Phase
Reversed Phase
Phenyl-Hexyl
HPLC
Anti-Bacterial Agents

Applications

Immunosuppressants by LCMS using Luna 5um Phenyl-Hexyl 50x2.1mm
Analytes
1Everolimus
2Sirolimus
3Tacrolimus
4Ascomycin
5Cyclosporin D
Compound Name
Everolimus
CID
6442177
Molecular Formula
C53H83NO14
Molecular Weight
958
No. Hydrogen Bond Acceptors
14
No. Hydrogen Bond Donors
3
Details
LC Conditions (App ID: 20688)
Column:
Brand Name:
Part No:
Phase Name:
Phenyl-Hexyl
Elution Type:
Isocratic
Mobile Phase:
A: 97% methanol + 0.3 mM ammonium acetate + 0.003% acetic acid
Gradient Profile:
Step No.Time(min)%A
10100
Flow Rate:
0.5 mL/min
Temperature:
22°C
Detection:
Tandem Mass Spec (MS-MS) (22°C)
Detector Info:
SCIEX API 3000™

Order Items Used in This Application

Similar Applications
search
No Data