FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
TCA's LC/MS/MS Kinetex 1.7µm C18 50x2.1mm Agilent 1200
19307
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Antidepressive Agents, Tricyclic
Clinical

Applications

TCA's LC/MS/MS Kinetex 1.7µm C18 50x2.1mm Agilent 1200
Analytes
1Doxepin
2Desipramine
3Imipramine
4Nortriptyline
5Amitriptyline
6Trimipramine
Compound Name
Doxepin
CID
667477
Molecular Formula
C19H21NO
Molecular Weight
279
No. Hydrogen Bond Acceptors
2
Details
LC Conditions (App ID: 19307)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: Water with 0.05% formic acid
B: Acetonitrile with 0.05% formic acid
Gradient Profile:
Step No.Time(min)%A%B
107030
226040
33595
43.17030
54.57030
Flow Rate:
0.6 mL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
SCIEX API 4000™

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm PS C18 100 Å, LC Column 100 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Kinetex 1.7 µm C18 100 Å, LC Column 50 x 2.1 mm, Ea
Reversed Phase
Luna 3 µm C18(2) 100 Å, LC Column 20 x 2 mm, Ea
Reversed Phase