For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Extraction of Benzodiazepines Using Strata Screen-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
18849
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Antiemetics
Forensics/Toxicology
Clinical
Toxicology / Drug Screening / TDM
Therapeutic / Clinical

Applications

Extraction of Benzodiazepines Using Strata Screen-C on Kinetex 2.6µm C18 50 x 2.1mm ID by LCMS
Analytes
1alpha-Hydroxy alprazolam
2alpha-HydroxyAlprazolam-d5
3Oxazepam-d5
4Oxazepam
5Nordiazepam-d5
6Nordiazepam
7Lorazepam
8Clonazepam
9Temazepam-d5
10Temazepam
11Diazepam
Compound Name
Alpha-Hydroxyalprazolam
CID
162244
Molecular Formula
C17H13ClN4O
Molecular Weight
324
No. Hydrogen Bond Acceptors
4
No. Hydrogen Bond Donors
1
Details
LC Conditions (App ID: 18849)
Column:
Brand Name:
Kinetex
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1%FA in H2O
B: 0.1%FA in Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
106535
226535
32.01595
42.3595
52.316535
64.16535
Flow Rate:
450 µL/min
Temperature:
0°C
Detection:
Mass Spectrometer (MS) (22°C)
Detector Info:
ESI conditions Positive Ionization NEB: 12.00 CUR: 11.00 IS: 4500.00 TEM: 450.00 CAD: 11.00 EP 10.00

Order Items Used in This Application

Similar Applications
search
No Data