FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
Interferon Alpha digests reduced & non-red on Jupiter 3u C18
18072
Separation Mode: Reversed Phase
Reversed Phase
C18
HPLC
Pharmaceutical/Biopharmaceutical
Life Science
Clinical Research
Therapeutic / Clinical

Applications

Interferon Alpha digests reduced & non-red on Jupiter 3u C18
Analytes
1Reduced Interferon
2Interferon

Analyte not found

Details
LC Conditions (App ID: 18072)
Column:
Brand Name:
Part No:
Phase Name:
C18
Elution Type:
Gradient
Mobile Phase:
A: 0.1% TFA and 2% Acetonitrile in Water
B: 0.085% TFA, 90% Acetonitrile in Water
Gradient Profile:
Step No.Time(min)%A%B
10991
2355050
3402080
Flow Rate:
0.3 mL/min
Temperature:
25°C
Detection:
UV-Vis Abs.-Diode Array (PDA) @ 214 nm (25°C)
Sample Note:
Application Focus: Using peptide mapping with Jupiter 300 3u C18 to identify disulfide bonds Peptide mapping using different proteases is a good analytical method to screen PTMs of therapeutic proteins and is shown on Jupiter 300 3u C18 for a biogeneric alpha interferon. Interferon-alpha is a 19.2 kDa protein containing two disulfide bonds (cys1-cys98; cys29-cys138) and was tryptic digested (1:20 E/S ratio, 18 hrs at 37C) and then half of the digest was reduced with DTT (3 mM DTT at 45°C for 30 min). Both native and reduced tryptic digests injected on the Jupiter 300 C18 columns and overlaid. App ID 18072 shows that peptides retention time shifting due to disulphide bonds breakage after reduction reaction. Such shifts are clearly observed in the peak distribution at different retention times of the trypsin digested Interferon- alpha(lower chromatogram) and Interferon-alpha digested and reduced (upper chromatogram). Such shifts are indicative of a disulfide containing peptide and are easily verified by mass spectrometry. The high resolution that the Jupiter 300 3u C18 column delivers makes it an excellent choice for peptide mapping.