For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
ASTM D5501: Determination of Denatured Bioethanol on ZB-Bioethanol
16667
Separation Mode: Application Specific
Application Specific
ASTM D 5501
ZB-Bioethanol
GC
Anti-Infective Agents, Local
Fuels
Alternative Fuel
Energy / Petroleum / Plastic

Applications

ASTM D5501: Determination of Denatured Bioethanol on ZB-Bioethanol
Analytes
1Methanol
2Ethanol
Compound Name
Methanol
CID
887
Molecular Formula
CH4O
Molecular Weight
32
No. Hydrogen Bond Acceptors
1
No. Hydrogen Bond Donors
1
Details
Official Method: ASTM D 5501
GC Conditions (App ID: 16667)
Column:
Part No:
Phase Name:
ZB-Bioethanol
Injection:
Split 50:1 @ 300°C, 1.0 μL
Oven Program:
55 °C for 1.7 min to 260 °C @ 40 °C/min for 2.67 min.
Carrier Gas:
Constant Flow Helium, 25.0 cm/sec
Detection:
Flame Ionization (FID) (300°C)
Sample Notes:
Instrumentation: Shimadzu GC-2010

Order Items Used in This Application

Similar Applications
search
Zebron ZB-DHA-PONA, GC Cap. Column 50m x 0.20mm x 0.5µm, Ea
Application Specific
Zebron ZB-624, GC Cap. Column 30 m x 0.25 mm x 1.40 µm, Ea
Standard
Zebron ZB-FFAP, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-FFAP, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-FFAP, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-Bioethanol, GC Cap. Column 15 m x 0.25 mm x 1.00 µm, Ea
Application Specific
Zebron ZB-Bioethanol, GC Cap. Column 30 m x 0.25 mm x 1.00 µm, Ea
Application Specific
Zebron ZB-Bioethanol, GC Cap. Column 30 m x 0.25 mm x 1.00 µm, Ea
Application Specific
Zebron ZB-1, GC Cap. Column 100 m x 0.25 mm x 0.50 µm, Ea
Standard
Zebron ZB-1, GC Cap. Column 100 m x 0.25 mm x 0.50 µm, Ea
Standard