For security purposes
FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Applications
EPA Method 553, Benzidines and Nitrogen-Containing Pesticides
14473
Separation Mode: Reversed Phase
Reversed Phase
EPA 553
C18(2)
HPLC
Uncoupling Agents
Environmental
Pollutant / Contaminant
Environmental / Health

Applications

EPA Method 553, Benzidines and Nitrogen-Containing Pesticides
Analytes
1Benzidine
2Benzoylprop ethyl
3Caffeine
4Carbaryl
5o-Chlorophenyl thiourea
63,3'-Dichlorobenzidine
73,3'-Dimethoxybenzidine
83,3'-Dimethylbenzidine
9Diuron
10Linuron
11Monuron
12Rotenone
13Siduron
Compound Name
Benzidine
CID
7111
Molecular Formula
C12H12N2
Molecular Weight
184
No. Hydrogen Bond Acceptors
2
No. Hydrogen Bond Donors
2
Details
Official Method: EPA 553
LC Conditions (App ID: 14473)
Column:
Brand Name:
Part No:
Phase Name:
C18(2)
Elution Type:
Gradient
Mobile Phase:
A: 10mM Ammonium Acetate:Acetonitrile (75:25)
B: Acetonitrile
Gradient Profile:
Step No.Time(min)%A%B
107525
217525
3294060
4494060
Flow Rate:
0.3 mL/min
Temperature:
22°C
Detection:
UV-Vis Abs.-Variable Wave.(UV) @ 210 nm (22°C)
Sample Note:
SecurityGuard™ Guard Cartridge System extends column lifetime. - SecurityGuard Cartridges, C18 4 x 3.0mm, 10/Pk Part No.: AJ0-4287 - Holder Part No.: KJ0-4282

Order Items Used in This Application

Similar Applications
search
Kinetex 2.6 µm EVO C18 100 Å, LC Column 150 x 2.1 mm, EA
Reversed Phase
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-SemiVolatiles, GC Cap. Column 20 m x 0.18 mm x 0.36 µm, Ea
Application Specific
Zebron ZB-XLB-HT, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
High Temperature
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-XLB, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Standard
Zebron ZB-MultiResidue-1, GC Cap. Column 30 m x 0.25 mm x 0.25 µm, Ea
Application Specific
Zebron ZB-5MS, GC Cap. Column 20 m x 0.18 mm x 0.32 µm, Ea
Standard