FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Product Information
Phase Detail
Strata C18-E SPE Solid Phase Extraction (SPE)
Elevate sample preparation with Phenomenex's Strata C18 Solid Phase Extraction products. Achieve superior chromatography results with Strata C18-E SPE sorbent.

Strata C18-E SPE Solid Phase Extraction (SPE)

Strong Hydrophobic Retention Solid Phase Extraction (SPE)

Strata C18-E is a silica-based end-capped C18 solid phase extraction (SPE) sorbent that offers strong hydrophobic retention with negligible secondary polar interactions from active silanol groups. Strata C18-E can be best used for the extraction of hydrophobic or polar organic analytes from aqueous matrices. In addition, Strata SPE sorbents are manufactured with the highest purity and extensively tested to give clean extracts.

Strata C18-E is compatible with various formats, including C18 SPE cartridges. Our C18 SPE products are synonymous with quality, offering the highest purity and undergo rigorous testing to deliver consistently clean extracts for your analytical needs.

Surface Area (m²/g)
500, m²/g
Separation Mode
Reversed Phase
Pore Size
70, 100
Particle Size
20, 55
Recommended Use
Extraction of hydrophobic molecules

Available Formats

strata-well-plate
96-Well-Plate

strata-3-tubes
1, 3, and 6mL Tubes

strata-3-giga-tubes
Giga Tubes (12-150mL Tubes)

mercury-cartridge-column
On-line Extraction Columns and Cartridges

Starting Method

Strata Tube

Strata Reversed Phase

Condition
1 mL Methanol
Equilibrate
1 mL DI Water
Load
Pretreated sample
Wash
1 mL 5% Methanol in DI Water, dry under vacuum for 2-5 min
Elute
1 mL Methanol
Based on 100 mg/1 mL sorbent mass. The above is a convenient starting point for SPE method development, further optimization may be required to tailor the method for your specific extraction needs.